Epte System percutaneous electrolysis equipment
2- Facilidades de pago
- 14-días para retornar
- Entrega: 1-2 Días
EPTEAN® SYSTEM Percutaneous Electrolysis Device for the treatment of tendinopathies with health CE certificate.
EPTE® briefcase includes:
- EPTE® equipment.
- 1 connection cable.
- 3 AA 1.5V batteries
does not include
Pedal-Puller.
the official EPTE® formation. Query
- 1 Needle holder.
- 1 needle kit
- Chlorhexidine
- 4 contact electrodes
EPTE® System, percutaneous electrolysis device with sanitary CE certificate.
Equipment designed specifically for the application of percutaneous electrolysis with security towards the patient as well as guarantees and tranquility for the professional.
It allows to apply an invasive physiotherapy technique for the treatment of tendinosis and fibrillar breakage, through the application of microcurrent through a acupuncture needle to access the soft tissue damaged and induce the self -rerection.
With the application of the EPTE® and, thanks to its intensity modulator (µA/m2), we produce an electrical stimulus that causes the decomposition of the degenerate soft tissue and causes an inflammatory response, necessary for any repair, without the healthy tissue See affected. Thanks to Epte, we can treat soft tissues and get excellent results. It is a pioneer technique, with painless and rapid recovery currents
Technical specifications:
General
Dimensions: 220x140x46mm.
Container: High impact resistance ABS box.
Weight: 350g (without batteries)
Food: 3 alkaline batteries of 1.5V non -rechargeable type AA.
Consumption: 35th.
Autonomy: 3100 treatments (500μA / 10kΩ for 80sg)
Classification of the Team: Class IIA according to rule 9 of Directive 93/42/CEE on health products.
Percutaneous electrolysis therapy
Current waveform: continuous [1].
Maximum output voltage: 48v.
Minimum output current: 50μA.
Maximum output current: 2500μA.
Minimum current resolution: 10μA.
Minimum time resolution: 5μSg.
Maximum treatment time: unlimited.
[1] The treatment is composed of two current modes, initially the current ascends until the nominal value of the treatment begins, where the stationary or continuous mode begins.
Screen
Dimensions: 80x36 mm.
Type: reflective alphanumeric LCD screen.
Resolution: 16 characters x 2 lines.
Maximum vertical viewing angle: 40 º
Maximum horizontal viewing angle: 30 º
Environmental specifications
Operating temperature: from 0 ºC to 40 ºC
Storage and transport temperature: -20ºC to 60 ºC
Humidity: 5% to 95% relative humidity, without condensation.
Electromagnetic compatibility test standards: in 60601-1-2 (2007). Corrigendum 2010.
Electrical security test standards: in 60601-1: 2006. Corrigendum 2010 IEC 60601-2-10: 2012.
To install this Web App in your iPhone/iPad press and then Add to Home Screen.